Portable

PRC 2X

  • Compact, Light Weight Unit with Holster
  • Capability to Measure & Simulate DC Voltages, DC Current,
  • Frequency, Resistance, Thermocouples & RTDs
  • Dual Display with White LED Back Light
  • Capability to Provide 24V Loop supply (24V) & Continuity Test
  • Simulate Two Wire Transmitter

Suitable for Calibration of Thermocouples, RTDs, Thermometers, Transmitters, Indicators, Controllers, Receivers, mA, V, mV, Ohms, Frequency Measuring Instruments etc.

Specifications

DC VOLTAGE / DC CURRENT/ FREQUENCY MEASURE & SOURCE

FunctionsRangeMaximum
measurement range
ResolutionAccuracy ( % of
reading + counts)
Remarks
DC Voltage
(Measure)
28V (upper display)-33V ~ 33V0.001V0.025 + 2-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Input resistance : >1MΩ.
200mV (upper display)-80mV ~ 80mV0.001mV0.025 + 20
-200mV ~ 200mV0.01mV0.025 + 2
50V (lower display)-1V ~ 60V0.001V0.02 + 2
100mV (lower display)
-15mV ~ 80mV0.001mV0.02 + 20
80mV ~ 125mV0.01mV0.02 + 2
DC Voltage
(Source)
100 mV-15mV ~ 99.999mV0.001mV0.02 + 20-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load : 1mA or 1kΩ (It should be based on the lower load.)
100mV ~ 125mV0.01mV0.02 + 2
10V0 ~ 11V0.001V0.02 + 2
DC Current mA
Measure)
20mA (upper display)-24mA ~ 24mA0.001mA0.025 + 2-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Input resistance < 100Ω.
20mA (lower display)0 ~ 24mA0.001mA0.025 + 2
DC Current mA
(Source)
20mA0 ~ 24mA0.001mA0.02 + 2-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load voltage : 20V, equivalent to voltage of 20mA on 1000Ω load resistance.
20mA (Transducer simulation)0 ~ 24mA0.001mA0.01 + 1
Frequency
(Measure)
100Hz1 ~ 99.999H1 ~ 99.999H0.01 + 1Sensitivity : 10Hz ~ 10kHz, Vp-p≥1V; rest : Vp-p≥2V. Wave form : Square wave. 5 counting points should be added to errors of other wave forms. Commercial frequency can be measured directly.
1000Hz100 ~ 999.99Hz0.01Hz0.01 + 1
10kHz1k ~ 9.9999kHz0.1Hz0.01 + 1
100kHz10k ~ 99.999kHz1Hz0.01 + 1
Frequency
(Source)
200Hz0 ~ 200Hz0.01Hz0.01 + 1Output amplitude : ≥4.5Vp-p; Wave form : Square wave
2000Hz0 ~ 2000Hz0.1Hz0.01 + 1
20kHz0 ~ 20kHz1Hz0.01 + 1

Technical Datasheet

RESISTANCE MEASURE & SOURCE Functions Range Maxim

FunctionsRangeMaximum
measurement range
ResolutionAccuracy ( % of
reading + counts)
Remarks
Resistance (Ohm) – UPPER display
(Measure)
400Ω0 ~ 440Ω0.01Ω0.05%rdg + 2 count-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load voltage : 20V, equivalent to voltage of 20mA on 1000Ω load resistance.
3200Ω420Ω ~ 3300Ω0.05%rdg + 2 count
On-off test0 ~ 200Ω0.05%rdg + 1 count
Resistance (Ohm) – Lower display
(Measure)
400Ω0 ~ 440Ω0.01Ω0.15 Ω (2 wire & 0.10 Ω (4 wire)-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Exciting current during measurement : 400Ω:0.4mA±10%; 3200Ω:0.2mA±10%。 Two wire : Does not include lead resistance. Three wire : Assumes matched leads with a total resistance not exceeding 25Ω.
3200Ω420Ω ~ 3600Ω0.1Ω1.0 Ω (2 wire & 0.5 Ω (4 wire)
Resistance
(Source)
400Ω (External exciting current : 0.40mA ~ 3.30mA )0 ~ 440Ω0.01Ω0.15 Ω-10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C.
3200Ω(External exciting current : 0.1mA ~ 0.6mA400Ω ~ 3600Ω0.1Ω0.50 Ω

THERMOCOUPLE & RTD – MEASURE & SOURCE

FunctionsRangeResolutionAccuracy °CRemarks
J-200°C ~ 0°C
0°C ~ 1200°C
0.1°C1.0°C
0.7°C
Errors of cold-junction compensation are not included in the table. Accuracy of cold-junction compensation : 1.5°C
K-200°C ~ 0°C
0°C ~ 1370°C
1.2°C0.8°C
1.2°C
T-200°C ~ 0°C
0°C ~ 400°C
0.1°C1.2°C
0.8°C
E-200°C ~ 0°C
0°C ~ 950°C
0.1°C1.2°C
0.8°C
R-20°C ~ 0°C
0°C ~ 500°C
500°C ~ 1750°C
0.1°C2.5°C
1.8°C
1.4°C
S-20°C ~ 0°C
0°C ~ 500°C
500°C ~ 1750°C
0.1°C2.5°C
1.8°C
1.5°C
B600°C ~ 800°C
800°C ~ 1000°
1000°C ~ 180
0.1°C2.2°C
1.8°C
1.4°C
N-200°C ~ 0°C
0°C ~ 1300°C
0.1°C1.5°C
0.9°C
Pt100-200°C ~ 850°c0.1°C2 / 3 wire : 0.4
4 wire : 0.3
Output : 0.3
As for exciting current during measurement, please refer to resistance measurement function. As for allowable external exciting current during output, please refer to resistance output function. 2 wire : Does not include lead resistance. 3 wire : Assumes matched leads with a total resistance not exceeding 25Ω.
Pt1000-200°C ~ 650°C0.1°C2 / 3 wire : 0.3
4 wire : 0.15
Output : 0.15
Cu50-50°C ~ 150°0.1°C2 / 3 wire : 0.8
4 wire : 0.5
Output : 0.5
Cu100-50°C ~ 150°C0.1°C2 / 3 wire : 0.4
4 wire : 0.25
Output : 0.25
  • Multifunction Process Calibrator
  • 1.2Vdc “AAA” Rechargeable Battery (6 Nos.) with Power Adaptor (12V / 1A)
  • Operation Manual
  • Test Leads & Alligator Clips – 2sets
  • Carrying Case
  • Traceable Calibration Certificate