Portable/Handheld

Nagman® 02+

  • Compact, Light Weight Unit with Holster
  • Selectable Temperature Output Readings (in °C / °F)
  • Measurement & Simulation of  Thermocouple types R, S, B, E, K, J, T & N

Input Functions (Measurement)

Input Range Input range Resolution Accuracy Remarks
Voltage (DC) 100 mV -10 to 110 mV 10 mV ±0.05% measured value

± 30 mV

Input Resistance 1MW
Thermocouple R -40 to 1760°C 1°C ±0.05% measured value ±3°C (<100°C)

±0.05% measured value ±2°C (>100°C)

Input Resistance 1MW
S -20 to 1760°C 1°C
B 400 to 1800°C 1°C ±0.05% measured value ±3°C (400 to 600°C)

±0.05% measured value ±2°C (>600°C)

E -200 to 1000°C 0.1°C ±0.05% measured value ±2°C (< -100°C)

±0.05% measured value ±1°C (> -100°C)

K -200 to 1370°C 0.1°C
J -200 to 1200°C 0.1°C
T -200 to 400°C 0.1°C
N -200 to 1300°C 0.1°C

Output Functions (Simulation)

Output Range Output range Resolution Accuracy Remarks
Voltage (DC) 100 mV (Low) -10.00 to 110.00 mV 0.01 mV ±0.05% set value ±30 mV Max. output current ±2 mA
1V (High) -0.10 to 1.10 V 0.1 mV ±0.05% set value ±0.3 mV
Thermocouple R -40 to 1760°C 1°C ±0.05% set value ±3°C (<100°C)

±0.05% set value ±2°C (>100°C)

By using ITS-90 temperature scale
S -20 to 1760°C 1°C
B 400 to 1800°C 1°C ±0.05% set value ±3°C (400 to 600°C)

±0.05% set value ±2°C (>600°C)

E -200 to 1000°C 0.1°C ±0.05% set value ±2°C (<100°C)

±0.05% set value ±1°C (>100°C)

K -200 to 1370°C 0.1°C
J -200 to 1200°C 0.1°C
T -200 to 400°C 0.1°C
N -200 to 1300°C 0.1°C
  • Calibration Certificate from NABL Accredited – as per ISO/IEC 17025:2005 Laboratory

Calibration Certificates are issued in Accordance with our Scope as granted by NABL per ISO/IEC 17025:2017 Standards