Portable

PRC 5X

  • Compact, Light Weight Unit with Holster
  • Capability to Measure & Simulate DC Voltages, DC Current,
  • Frequency, Resistance, Thermocouples & RTDs
  • Dual Display with White LED Back Light
  • Capability to Provide 24V Loop supply (24V) & Continuity Test
  • Simulate Two Wire Transmitter

Suitable for Calibration of Thermocouples, RTDs, Thermometers, Transmitters, Indicators, Controllers, Receivers, mA, V, mV, Ohms, Frequency Measuring Instruments etc.

Specifications

DC VOLTAGE / DC CURRENT/ FREQUENCY MEASURE & SOURCE

Functions Range Maximum
measurement range
Resolution Accuracy ( % of
reading + counts)
Remarks
DC Voltage
(Measure)
28V (upper display) -33V ~ 33V 0.001V 0.05 + 2 -10°C ~ 18°C, + 28°C ~ 55°C
temperature coefficient, ±0.005%FS / °C. Input resistance : >1MΩ.
200mV (upper display) -80mV ~ 80mV 0.001mV 0.05 + 20
-200mV ~ 200mV 0.01mV 0.05 + 2
50V (lower display) -1V ~ 60V 0.001V 0.05 + 2
100mV (lower display)
-15mV ~ 80mV 0.001mV 0.05 + 20
80mV ~ 125mV 0.01mV 0.05 + 2
DC Voltage
(Source)
100 mV -15mV ~ 99.999mV 0.001mV 0.05 + 20 -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load : 1mA or 1kΩ (It should be based on the lower load.)
100mV ~ 125mV 0.01mV 0.05 + 2
10V 0 ~ 11V 0.001V 0.05 + 2
DC Current mA
Measure)
20mA (upper display) -24mA ~ 24mA 0.001mA 0.05 + 2 -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Input resistance < 100Ω.
20mA (lower display) 0 ~ 24mA 0.001mA 0.05 + 2
DC Current mA
(Source)
20mA 0 ~ 24mA 0.001mA 0.05 + 2 -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load voltage : 20V, equivalent to voltage of 20mA on 1000Ω load resistance.
20mA (Transducer simulation) 0 ~ 24mA 0.001mA 0.05 + 2
Frequency
(Measure)
100Hz 1 ~ 99.999H 1 ~ 99.999H 0.02 + 1 Sensitivity : 10Hz ~ 10kHz, Vp-p≥1V; rest : Vp-p≥2V. Wave form : Square wave. 5 counting points should be added to errors of other wave forms. Commercial frequency can be measured directly.
1000Hz 100 ~ 999.99Hz 0.01Hz 0.02 + 1
10kHz 1k ~ 9.9999kHz 0.1Hz 0.02 + 1
100kHz 10k ~ 99.999kHz 1Hz 0.02 + 1
Frequency
(Source)
200Hz 0 ~ 200Hz 0.01Hz 0.02 + 1 Output amplitude : ≥4.5Vp-p; Wave form : Square wave
2000Hz 0 ~ 2000Hz 0.1Hz 0.02 + 1
20kHz 0 ~ 20kHz 1Hz 0.02 + 1

RESISTANCE MEASURE & SOURCE Functions Range Maxim

Functions Range Maximum
measurement range
Resolution Accuracy ( % of
reading + counts)
Remarks
Resistance (Ohm) – UPPER display
(Measure)
400Ω 0 ~ 440Ω 0.01Ω >0.25 Ω (2 wire & 0.15 Ω (4 wire) -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Maximum load voltage : 20V, equivalent to voltage of 20mA on 1000Ω load resistance.
3200Ω 420Ω ~ 3300Ω 0.05%rdg + 2 count
On-off test 0 ~ 200Ω 0.05%rdg + 1 count
Resistance (Ohm) – Lower display
(Measure)
400Ω 0 ~ 440Ω 0.01Ω 0.25 Ω (2 wire & 0.15 Ω (4 wire) -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C. Exciting current during measurement : 400Ω:0.4mA±10%; 3200Ω:0.2mA±10%。 Two wire : Does not include lead resistance. Three wire : Assumes matched leads with a total resistance not exceeding 25Ω.
3200Ω 420Ω ~ 3600Ω 0.1Ω 1.5 Ω (2 wire & 1.0 Ω (4 wire)
Resistance
(Source)
400Ω (External exciting current : 0.40mA ~ 3.30mA ) 0 ~ 440Ω 0.01Ω 0.25 Ω -10°C ~ 18°C, + 28°C ~ 55°C temperature coefficient, ±0.005%FS / °C.
3200Ω(External exciting current : 0.1mA ~ 0.6mA 400Ω ~ 3600Ω 0.1Ω 1.0 Ω

THERMOCOUPLE & RTD – MEASURE & SOURCE

Functions Range Resolution Accuracy °C Remarks
J -200°C ~ 0°C
0°C ~ 1200°C
0.1°C 1.5°C
1.0°C
Errors of cold-junction compensation are not included in the table. Accuracy of cold-junction compensation : 1.5°C
K -200°C ~ 0°C
0°C ~ 1370°C
0.1°C 1.8°C
1.2°C
T -200°C ~ 0°C
0°C ~ 400°C
0.1°C 1.8°C
1.2°C
E -200°C ~ 0°C
0°C ~ 950°C
0.1°C 1.5°C
1.0°C
R -20°C ~ 0°C
0°C ~ 500°C
500°C ~ 1750°C
0.1°C 4°C
2.5°C
2°C
S -20°C ~ 0°C
0°C ~ 500°C
500°C ~ 1750°C
0.1°C 4°C
2.5°C
2°C
B 600°C ~ 800°C
800°C ~ 1000°
1000°C ~ 180
0.1°C 3.5°C
2.5°C
2°C
N -200°C ~ 0°C
0°C ~ 1300°C
0.1°C 2.0°C
1.2°C
Pt100 -200°C ~ 850°c 0.1°C 2 / 3 wire : 0.7
4 wire : 0.4
Output : 0.7
As for exciting current during measurement, please refer to resistance measurement function. As for allowable external exciting current during output, please refer to resistance output function. 2 wire : Does not include lead resistance. 3 wire : Assumes matched leads with a total resistance not exceeding 25Ω.
Pt1000 -200°C ~ 650°C 0.1°C 2 / 3 wire : 0.4
4 wire : 0.3
Output : 0.3
Cu50 -50°C ~ 150° 0.1°C 2 / 3 wire : 1.2
4 wire : 0.8
Output : 0.8
Cu100 -50°C ~ 150°C 0.1°C 2 / 3 wire : 0.7
4 wire : 0.4
Output : 0.4
  • Multifunction Process Calibrator
  • 1.2Vdc “AAA” Rechargeable Battery (6 Nos.) with Power Adaptor (12V / 1A)
  • Operation Manual
  • Test Leads & Alligator Clips – 2sets
  • Carrying Case
  • Traceable Calibration Certificate
  • Calibration Certificates are issued in Accordance with our Scope as granted by NABL per ISO/IEC 17025:2017
    Standards

Click here to download our product catalog

Other Products
03-Big

Nagman 02+

View Product

03-Big

Nagman 03+

View Product

05-Big

Nagman 05+

View Product

© Nagman 2024 — All rights reserved.