Portable/Handheld
Nagman® 15+
- Compact, Light Weight Unit with Holster
- Capability to Measure & Simulate DC Voltages, Resistances, Thermocouples and RTD
- 5 digits display of Measurement & 6 digits display for Simulation
- Dual LCD with white LED for backlight
- Capability to perform Continuity Test & provide Loop Supply (24 V) for Transmitters
- Measurement & Simulation of :
- DC Current : 50 mA in Measurement Mode; 20 mA in Simulation Mode
- RTD Types : PT100, Pt200, Pt500, Pt1000, Cu10 & Cu50
- Frequency : 50 Hz, 5 KHz & 50 KHz in Measurement Mode, 100 KHz, 1 KHz & 10 KHz in Simulation Mode
- DC Voltages : 200 mV, 5 V & 50 V in Measurement Mode, 1000 mV & 10 V in Simulation Mode
Input Functions (Measurement)
Functions | Reference | Ranges | Resolution | Accuracy | Remarks |
DC Voltage | 200 mV | -20 to 220 mV | 10 mV | 0.02 + 0.02 | Impedance approximately 100 MΩ |
5 V | -0.5 to 5.5 V | 0.1 mV | 0.02 + 0.01 | Impedance approximately 1 MΩ | |
50V | -5 to 55 V | 1 mV | 0.03 + 0.01 | ||
DC mA | 50 mA | -4 to 55 mA | 1 mA | 0.02 + 0.01 | Impedance : 5 Ω |
Frequency | 500 Hz | 3 to 500 Hz | 0.01 Hz | ±2 digit | Impedance : 100 kΩ Min. Sensitivity : 3 Vp-p min. Duty Cycle : 50% |
5 KHz | 3 Hz to 5 KHz | 0.1 Hz | |||
50 KHz | 3 Hz to 50 KHz | 1 Hz | |||
Continuity | <250 Ω | Close / open | — | — | Open Circuit Voltage : about 2.5 V |
Output Functions (Simulation)
Functions | Reference | Ranges | Resolution | Accuracy | Remarks |
DC Voltage | 1000 mV | -100 to 1100 mV | 10 mV | 0.02 + 0.01 | Maximum output : 2 mA |
10 V | -1 to 11 V | 0.1 mV | 0.02 + 0.01 | Maximum output : 5 mA | |
DC Current | 20 mA | 0 to 22 mA | 1 mA | 0.02 + 0.02 | Simulator Transmitter : 5 to 28 V Power Supply outside; 1 KΩ at 20 mA |
Frequency | 100 Hz | 1 to 110 Hz | 0.01 Hz | ±2 count | Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ Duty Cycle : 50% |
1 KHz | 0.1 to 1.1 KHz | 1 Hz | |||
10 KHz | 1 to 11 KHz | 0.1 KHz | |||
Pulse | 100 Hz | 1 to 100000 cycles | 1 cycle | — | |
1 KHz | |||||
10 KHz | |||||
Loop | 24 V | — | — | ±10% | Maximum output current : 25 mA Short circuit protected |
* Accuracy is expressed as ±(percentage of Reading + percentage of Range)
- Voltage / Milliamp Calibrator – Model Nagman 15+
- 5V x 4 Nos. AAA Alkaline Batteries
- Operation Manual
- Test Leads
- Traceable Calibration Certificate
- Calibration Certificate from NABL Accredited – as per ISO/IEC 17025:2005 Laboratory